FRAS Smart characteristics and operating conditions:
| General Characteristics |
| * Power line: |
1 phase |
| * Power supply: |
230Vac, 50/60 Hz (on request 115Vac) |
| * Consumption: |
50 VA |
| Operating Conditions |
| * Temperature: |
15 : 32 °C (operating) 0 : 50°C (non operating) |
| * Relative humidity: |
20 : 80 % (operating) 0 : 90% (non operating) |
| * Altitude: |
< 2000m |
| * Pollution degree: |
2 |
| * Sonic Pressure: |
< 85dBA |
| European Directives EEC |
| * Safety: |
Directive 73/23 & 93/68 |
| * Electromagnetic compatibility: |
Directive 89/336 |
| * IVD: |
Directive 98/79 |
| Mechanical Characteristics |
| * Dimensions: |
25 * 20 * 12 cm |
| * Weight: |
1.2 kg |
Technical characteristics:
| Photometric system |
| * Light source: |
Longlife-monochromatic |
| * Spectral range: |
505 nm interferential filter with 8 nm bandwidth |
| * Photo detector: |
Solid state |
| * Measuring principle: |
Lambert Beer’s law |
| * Temperature: |
37°C |
| Interface |
| * Keyboard: |
5 keys: ENTER UP DOWN ESC MENU |
| * Display: |
2 lines da 16 characters, backlighted |
| * Connection: |
RS232 with 9 pin connector |
|